Reported Jan 13, 2016 · Terminated

Remote Patient Monitoring System, MDP 2000 Series, Part Numbers: MDP2040-0100, PGM340R3, or PGM340R4. The MDP2000 is…

Vidco, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Vidco, Inc. · 6175 Sw 112th Ave · Beaverton, OR · 97008-4838

Is this the product you have?

  • Remote Patient Monitoring System, MDP 2000 Series
  • Part NumbersMDP2040-0100, PGM340R3, or PGM340R4. The MDP2000 is intended for use in conjunction with patient monitoring equipment. The MDP2000 is not patient connected Examples of areas where the MDP200 is used include ICU, CCU, PACU, Emergency, Telemetry Step-down and other areas where patient's physiological information are to be observed at one or more locations. Typical information displayed includes ECG and blood pressure waveforms and numeric values which as heart rate and systolic, mean and diastolic pressure. This device acts as the Central Monitor Station, it permits from one (1) to sixteen (16) bedside monitor(s) to be connected simultaneously
Lot codes and dates
  • Serial numbers
    • 6533-6537
    • 6561
    • 6566-6581
    • 6539
    • 6540
    • 6558-6560
    • 6562
    • 6508-6528
    • 6563-6565
    • 6529
    • 6531
    • 6532
    • 6541
    • 6551
    • 6556
    • and 6557
Amount recalled
59 units

Where it was sold

Why it was recalled

Testing at customer site showed unit Remote Patient Monitoring System MDP2040-0100 in a continuous trap condition, not allowing system to reset and reboot. Two customers recently complained of the system freezing and it could only be restarted if the user re-applied power.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Vidco, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 16, 2016 · Data as of Sep 19, 5:20 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.