Reported Dec 30, 2020 · Terminated
Flexor Check-Flo Introducer, RPN KCFW-7.0-35-55-RB-HFANL1-HC, Order Number G49043 Flexor Introducers and Guiding…
Cook Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Cook Inc. · 750 N Daniels Way · Bloomington, IN · 47404-9120
Is this the product you have?
- Flexor Check-Flo Introducer, RPN KCFW-7.0-35-55-RB-HFANL1-HC, Order Number G49043 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature.
- Lot codes and dates
- Lot Numbers
85 of 85 codes
131920551319206713200423132004241320296113202963132053091320531513205323132053241322383313229686132296871323180313234792132348011323765413243208132432091324654413246548132494511324945213249453132546171325461813262329132623301327908513281530132815311328179013281791132866451328664613286647132866481328665013286651132866521330335813303361133033621330336613303367133033681330337113303372133033731330337413303375133063351330633613319484133194891331949513346087133490061335244413354260
- Lot Numbers
- Amount recalled
- 3797
This recall covers 21 products
The agency filed one notice per product. You are reading one of them; the other 20 are below.
8 more are in the archive.
Where it was sold
Why it was recalled
Stop using this product now
There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Cook Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.