Reported Jan 2, 2013 · Terminated

Access Immunoassay System, Access 2 Immunoassay System, Synchron LXi 725 Clinical System, UniCel DxC 600i Synchron…

Beckman Coulter Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Beckman Coulter Inc. · 250 S Kraemer Blvd · Brea, CA · 92821-6232

Is this the product you have?

    • Access Immunoassay System
    • Access 2 Immunoassay System
    • Synchron LXi 725 Clinical System
    • UniCel DxC 600i Synchron
  • Access Clinical System Part NumbersAccess  81600; Access 2  81600N, 386220, A69186, A25656, A12900; Access 2 (LXi)  A15642; Access 2 Section, DxC 600i  A25640. The Access Immunoassay System is an in vitro diagnostic device used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. The Synchron LXi and the UniCel DxC 600i are Access systems integrated with a chemistry analyzer
Lot codes and dates
  • Part Numbers
    • Access  81600
    • Access 2  81600N
    • 386220
    • A69186
    • A25656
    • A12900
    • Access 2 (LXi)  A15642
    • Access 2 Section
    • DxC 600i  A25640
Amount recalled
6,858 units

This recall covers 2 products

The agency filed one notice per product. You are reading one of them; the other one is below.

Where it was sold

  • Nationwide

Why it was recalled

It has been determined that, in some cases, pulley drive components have broken parts or have become loose, mispositioned, or disconnected from their drive shafts. When these defects occur, the pulley may not turn at all, or may turn at an incorrect speed, resulting in inadequate washing or mixing of the contents of the reaction vessel.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Beckman Coulter Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Dec 23, 2013 · Data as of Sep 19, 6:50 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.