Reported Nov 30, 2016 · Terminated

Greatbatch Stiffer Coaxial Micro-Introducer. Model #/ Name: 10904-001 Stiffer Coaxial Micro-Introducer 4FR; 10904-002…

Greatbatch Medical

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Greatbatch Medical · 2300 Berkshire Ln N · Plymouth, MN · 55441-4575

Is this the product you have?

  • Greatbatch Stiffer Coaxial Micro-Introducer. Model #/
  • Name10904-001 Stiffer Coaxial Micro-Introducer 4FR; 10904-002 Stiffer Coaxial Micro-Introducer 5FR. The radiopaque Stiffer Coaxial Micro-Introducer (Figure 1) consists of a dilator and sheath, available in 4FR and 5FR. The inner dilator includes a stainless steel stiffening hypotube. The Coaxial Micro-Introducer is indicated for percutaneous introduction of up to a 0.038 in. guidewire or catheter into the vascular system through an initial puncture of a 21G introducer. The device is sold to customers in a bulk non-sterile configuration
Lot codes and dates
    • GBP3176925
    • GBP3176927
    • VN3470927
    • VN3470929
    • VN3486000
    • VN3485998
    • VN3495047
    • GBP3421664
    • VN3433259
    • VN3451654
    • VN3487654
    • VN3491069
    • VN3470925
    • VN3451646
Amount recalled
23,075

Where it was sold

Why it was recalled

Greatbatch Medical has identified an issue with the Stiffer Coaxial Micro-Introducer 4FR and 5FR after a field complaint reported that the rotating luer detached from the dilator hub during use.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Greatbatch Medical recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Sep 8, 2017 · Data as of Sep 19, 5:32 PM UTC