Reported Jan 20, 2016 · Terminated

FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.

Becton, Dickinson and Company, BD Biosciences

Low riskThis product is unlikely to cause harm, but it breaks a safety or labelling rule.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Becton, Dickinson and Company, BD Biosciences · 2350 Qume Dr · San Jose, CA · 95131-1812

Is this the product you have?

    • FMC7 FITC
    • Catalog No. 340918 Analyte Specific Reagent
Lot codes and dates
    • Lot No. 4162520
    • 4206855
    • 4267919
    • 4276818
    • 4309977
    • 5009530
    • 5036624
    • 5085573
    • 5182818
Amount recalled
1,388 units

Where it was sold

  • Nationwide

Why it was recalled

Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately 87%.

This product is unlikely to cause harm, but it breaks a safety or labelling rule.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Becton, Dickinson and Company, BD Biosciences recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Oct 31, 2016 · Data as of Sep 19, 6:02 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.