Reported Jan 9, 2013 · Terminated
ABX PENTRA AST CP, Part Number: A11A01629. The ABX Pentra AST CP reagent is a diagnostic reagent for quantitative in…
Horiba Instruments, Inc. dba Horiba Medical
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Horiba Instruments, Inc dba Horiba Medical · 34 Bunsen · Irvine, CA · 92618-4210
Is this the product you have?
- ABX PENTRA AST CP
- Part NumberA11A01629. The ABX Pentra AST CP reagent is a diagnostic reagent for quantitative in vitro determination of Aspartate Amino Transferase (AST) in serum or plasma by colorimetry
- Lot codes and dates
- Lot number00456-4108
- Amount recalled
- 66886 units
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Horiba Medical is recalling the ABX PENTRA AST CP because during the AST application validation on the Pentra C200 using an Audit Systems linearity kit, it was determined that two higher values for AST (expected at 800 U/L and 1000 U/L) were recorded at 280 and 199 U/L without alarm.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Horiba Instruments, Inc. dba Horiba Medical recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.