Reported Feb 23, 2022 · Ongoing
Veradius Unity, Model #718132
Philips Medical Systems Nederland B.v.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Veenpluis 4-6 · Best, N/A · N/A
Is this the product you have?
- Veradius Unity, Model #718132
- Lot codes and dates
- Wireless Foot switch 3P459801238191
- Wireless Foot switch 3P459801238231
- Amount recalled
- 12 US; 44 ROW
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
Why it was recalled
Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.