Reported Dec 18, 2019 · Terminated
Ligamax 5 mm Endoscopic Multiple Clip Applier
Ethicon Endo-Surgery Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ethicon Endo-Surgery Inc · 4545 Creek Rd · Blue Ash, OH · 45242-2803
Is this the product you have?
- Ligamax 5 mm Endoscopic Multiple Clip Applier
- Lot codes and dates
- Product Code11EL5ML
- Lot Numbers
T93X4RT93X9DT93Y2ZT93Y70T93Y96T93Z52T93Z76T9401CT9408FT94118T9416MT9422KT9426ZT94402T94536T94567T9461MT94667T94721T94758T94807T9484YT9487JT9491A T94A9KT94C2ZT94C8NT94D2NT94E66T94F3CT94F7FT94G99 T94H1NT94H78
- Amount recalled
- 48,428
Where it was sold
- Nationwide
- PR
Why it was recalled
A potential exists that certain LIGAMAX devices within the impacted lots may have small holes in the Tyvek lidding which could result in a breach of sterility.
This product could cause temporary or reversible health problems — stop using it.