Reported Feb 28, 2018 · Terminated

LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of…

Synthes, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Synthes, Inc. · 1302 Wrights Ln E · West Chester, PA · 19380-3417

Is this the product you have?

  • LCP Drill Sleeve 1.5 Drill Bit 1.1mm Part Number 03.114.001 Used in LCP Modular Mini Fragment System, for fixation of small bones and small bone fragments
Lot codes and dates
  • Lot numbers FT00227 9972971 H028400 H161606 H288454
Amount recalled
567 units

Why it was recalled

May have difficulty attaching the drill sleeve to the plate. If the drill sleeve will not connect to the plate intraoperatively, surgical delay could occur. If the pilot hole is drilled without the drill sleeve, the pilot hole could be off-center which could result in screw and/or plate loosening.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Synthes, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 25, 2018 · Data as of Sep 19, 6:01 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.