Reported Jan 2, 2019 · Terminated

Alinity i 25-OH Vitamin D Reagent Kit, List Number 08P4532

Abbott Ireland Diagnostics Division

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Abbott Ireland Diagnostics Division · Lisnamuck · Co. Longford · N/A

Is this the product you have?

    • Alinity i 25-OH Vitamin D Reagent Kit
    • List Number 08P4532
Lot codes and dates
  • UDI/Lot Numbers (01)00380740121952 (17)191011(10)92313UI00 / 92313UI00 (01)00380740121952 (17)190831(10)92484UI00 / 92484UI00
Amount recalled
760 units

This recall covers 6 products

The agency filed one notice per product. You are reading one of them; the other 5 are below.

Where it was sold

Why it was recalled

Abbott has become aware of a manufacturing issue that can result in damage to a bottle neck in some of the reagent cartridges found in the lots listed in Appendix A. If a bottle neck is damaged, it may cause a failure during the reagent loading process on the Alinity i instrument.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Abbott Ireland Diagnostics Division recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 1, 2021 · Data as of Sep 19, 4:46 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.