Reported Dec 25, 2019 · Terminated

A1CNow+ Systems (professional use product), Software Version Revision D, REF Numbers US: 3021, 3024; EU: 3037, 3038,…

Polymer Technology Systems

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Polymer Technology Systems · 4600 Anson Blvd · Whitestown, IN · 46075

Is this the product you have?

  • A1CNow+ Systems (professional use product), Software Version Revision D
  • REF Numbers US
    • 3021
    • 3024
  • EU
    • 3037
    • 3038
    • 3057
    • 3058
Lot codes and dates
  • Lot NumbersLot number

    112 of 112 codes

    • 1618030
    • 1630221
    • 1630210
    • 1606342
    • 1624606
    • 1611013
    • 1627740
    • 1618724
    • 1624608
    • 1625037
    • 1611132
    • 1619054
    • 1627111
    • 1619055
    • 1619056
    • 1624607
    • 1618817
    • 1602004
    • 1630111
    • 1619057
    • 1627240
    • 1611046
    • 1611015
    • 1618713
    • 1630211
    • 1611014
    • 1618031
    • 1625004
    • 1619032
    • 1627112
    • 1625003
    • 1624609
    • 1618735
    • 1606408
    • 1611141
    • 1625026
    • 1610954
    • 1611045
    • 1606331
    • 1703344
    • 1611131
    • 1602174
    • 1630013
    • 1618818
    • 1627718
    • 1602186
    • 1618736
    • 1611057
    • 1619033
    • 1610655
    • 1625036
    • 1602152
    • 1610932
    • 1606320
    • 1610931
    • 1610656
    • 1618839
    • 1610953
    • 1611056
    • 1624628
Amount recalled
112,648

This recall covers 5 products

The agency filed one notice per product. You are reading one of them; the other 4 are below.

Where it was sold

Why it was recalled

A numerical value less than 4% or greater than 13% may be displayed instead of the expected <4% or >13% result, due to a software bug.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Polymer Technology Systems recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jul 16, 2020 · Data as of Sep 19, 6:46 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.