Reported Mar 9, 2022 · Terminated

Titanium MPJ (Toe Implant) TOE JOINT TITANIUM PC MS, Part ID 17197

BioPro, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: BioPro, Inc. · 2929 Lapeer Rd · Port Huron, MI · 48060-2558

Is this the product you have?

  • Titanium MPJ (Toe Implant) TOE JOINT TITANIUM PC MS, Part ID 17197
Lot codes and dates
  • Lot Number128361
  • UDI00810012480277

This recall covers 10 products

The agency filed one notice per product. You are reading one of them; the other 9 are below.

Where it was sold

Why it was recalled

There is an incorrect statement regarding MR Compatibility in the BioPro Titanium MPJ implant IFU (IFU BP102303 rev 05) and patient implant cards (22266 rev 00) which indicates that the BioPro MPJ Implant has been tested for safety and compatibility in the MR Environment and is MR Conditional. MR Testing has not been conducted on the titanium version of the MPJ.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new BioPro, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 28, 2024 · Data as of Sep 19, 5:35 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.