Reported Feb 10, 2016 · Terminated
Emergency Transport Ventilator Oxylog 3000 Ventilator and Oxylog 3000 plus Ventilator Catalog Numbers: 2M86300/2M86965…
Draeger Medical, Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Draeger Medical, Inc. · 3135 Quarry Rd · Telford, PA · 18969-1042
Is this the product you have?
- Emergency Transport Ventilator Oxylog 3000 Ventilator and Oxylog 3000
- plus Ventilator Catalog Numbers2M86300/2M86965 and 5704813/5704831. Volume controlled and pressure-controlled emergency and transport ventilators
- Lot codes and dates
- All Oxylog 3000 Ventilators and Oxylog 3000 Plus Ventilators with the identified catalog numbers distributed between April 2007 - December 2015.
- Amount recalled
- 402 units (314 - Oxylog 3000; 88 - Oxylog 3000 Plus)
This recall covers 2 products
The agency filed one notice per product. You are reading one of them; the other one is below.
Where it was sold
Why it was recalled
Stop using this product now
Drager became aware of situations where the error message "Poti unplugged" was generated causing acoustic and visual alarms, the breathing system releases pressure and the ventilation function stops operating.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Draeger Medical, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.