Reported Mar 7, 2018 · Terminated
Coated VICRYL RAPIDETM (Polyglactin 910) Suture
Ethicon, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Ethicon, Inc. · Us Highway 22 West · Somerville, NJ · 08876
Is this the product you have?
- Coated VICRYL RAPIDETM (Polyglactin 910) Suture
- Lot codes and dates
- Coated VICRYL RAPIDETM (Polyglactin 910) Suture MPVR4260H Size 4-0 VICRYL RAPIDE Undyed 75cm PS-2 Needle 10705031149168 Coated VICRYL RAPIDETM (Polyglactin 910) Suture V2140H Size 4-0 VICRYL RAPIDE Undyed 70cm RB-1 Needle 10705031106048 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2217 Size 2-0 VICRYL RAPIDE Undyed 90cm V-34 Needle 10705031119925 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2253 Size 2-0 VICRYL RAPIDE Undyed 75cm FS-1 Needle 10705031119956 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2279 Size 4-0 VICRYL RAPIDE Undyed 75cm V-4 Needle 10705031086524 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2294 Size 4-0 VICRYL RAPIDE Undyed 75cm FS-3 Needle 10705031086548 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2297 Size 4-0 VICRYL RAPIDE Undyed 75cm FS-2 Needle 10705031119970 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR496 Size 4-0 VICRYL RAPIDE Undyed 18'' PS-2 Needle 10705031134690 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR945 Size 2-0 VICRYL RAPIDE Undyed 36'' CT-1 Needle 10705031133358 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VRI253H Size 2-0 VICRYL RAPIDE Undyed 75cm FS-1 Needle 10705031111172 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9941 Size 2-0 VICRYL RAPIDE Undyed 75cm FS Needle 10705031130999 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR426 Size 4-0 VICRYL RAPIDE Undyed 75cm PS-2 Needle 10705031134669 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR494 Size 4-0 VICRYL RAPIDE Undyed 45cm P-3 Needle 10705031134683 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR845 Size 4-0 VICRYL RAPIDE Undyed 45cm PC-3 Needle 10705031134744 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR917 Size 2-0 VICRYL RAPIDE Undyed 90cm V-34 Needle 10705031134768 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2286 Size 2-0 VICRYL RAPIDE Undyed 90cm FSL Needle 10705031080256 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VR2299 Size 2-0 VICRYL RAPIDE Undyed 75cm FS-2 Needle 10705031080294 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9918 Size 4-0 VICRYL RAPIDE Undyed 75cm PC-3 Needle 10705031151970 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9930 Size 4-0 VICRYL RAPIDE Undyed 75cm FS-2 Needle 10705031152007 Coated VICRYL RAPIDETM (Polyglactin 910) Suture FH1642H Size 4-0 VICRYL RAPIDE Undyed 45cm FS-2 Needle 10705031099968 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VRI214H Size 4-0 VICRYL RAPIDE Undyed 75cm RB-1 Plus Needle 10705031111127 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VRI312H Size 2-0 VICRYL RAPIDE Undyed 75cm SH-1 Plus Needle 10705031111264 Coated VICRYL RAPIDETM (Polyglactin 910) Suture VRI417H Size 2-0 VICRY RAPIDE Undyed 75cm SH Plus Needle 10705031111332 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9922 Size 4-0 VICRYL RAPIDE Undyed 75cm PS-2 Needle 10705031130845 Coated VICRYL RAPIDETM (Polyglactin 910) Suture W9933 Size 2-0 VICRYL RAPIDE Undyed 75cm PS Needle 10705031130937
- Amount recalled
- 129,372 Eaches (Internationally)
Why it was recalled
On August 17, 2017, Ethicon had decided to initiate a medical device recall (removal) of the twenty-five (25) lots of VICRYL RAPIDETM (Polyglactin 910) Suture that were distributed to the field. During finished goods testing, a suture damage was identified on a lot of VICRYL RAPIDETM Suture. During the course of the subsequent investigation, the suture also failed knot tensile strength testing.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Ethicon, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.