Reported Mar 7, 2018 · Terminated
A20975A "Working insert, with ramp, one way" (Lots 13ZW to 172W; A20976A "Bridge, one way" (Lots 146W to 172W) and…
Olympus Corporation of the Americas
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Olympus Corporation of the Americas · 3500 Corporate Pkwy · Center Valley, PA · 18034-0610
Is this the product you have?
- A20975A "Working insert, with ramp, one way" (Lots 13ZW to 172W; A20976A "Bridge, one way" (Lots 146W to 172W) and A20977A "Bridge, two way" (Lots 146W
- to 174W) Product UsageA20975A: Working insert for endoscopic diagnosis and treatment in urologic applications
- A20976ABridge for endoscopic diagnosis and treatment in urologic applications
- A20977ABridge for endoscopic diagnosis and treatment in urologic applications
- Lot codes and dates
- Model
- Numbers and Associated Lot NumbersA20975A -- 13-ZW to 172W A20976A -- 146W to 172W A20977A -- 146W to 174W
- Amount recalled
- 8741
Where it was sold
- Nationwide
Why it was recalled
Olympus has received complaints about fragments of adhesive which detached from inside the working channel of the referenced products. Cracking, chipping, missing pieces, and delamination of the adhesive have also been reported. In one case, a fragment of the adhesive was detected inside a patient's urethra during a procedure and was retrieved without injury to the patient. The manufacturer, Olympus Winter & lbe GmbH ("OWi") has determined that adhesive can be dislodged during the intended use of the cystoscopy bridge or working insert, e.g., when inserting an instrument through the working channel. As a result, a fragment of the adhesive may fall inside the patient's bladder, urethra or uterus and would need to be retrieved, extending the length of the procedure or requiring additional surgical treatment. The adhesive is used to seal gaps inside the working channel of the cystoscopy bridges and working inserts. In 2013, the adhesive used until then was discontinued and had to be replaced. Current investigations revealed that the durability of the new adhesive is insufficient
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.