Reported Jan 30, 2013 · Terminated
Abbott Laboratories, ARCHITECT 25-OH Vitamin D Reagent Kit, 03L52-25 (ROW) and 03L52-27 (U.S.) (1 x 100 tests),…
Abbott Laboratories
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Abbott Laboratories · 100 Abbott Park Rd · Abbott Park, IL · 60064-3502
Is this the product you have?
- Abbott Laboratories, ARCHITECT 25-OH Vitamin D Reagent Kit, 03L52-25 (ROW) and 03L52-27 (U.S.) (1 x 100 tests), 03L52-35 (ROW) and 03L52-37 (U.S.) (5 x 100 tests); in vitro diagnostic.
- Lot codes and dates
3 of 3 products
Products covered by this recall # Product as published 1 lot numbers 02412C000, exp. 12/26/2012; 02512C000, exp. 12/26/2012; 02012G000, exp. 7/16/2013; 01612G000, exp. 7/17/13; 00912G000, exp. 7/10/13; 02512I000, exp. 10/8/2013; 02612I000, exp. 10/8/2013; 03512I000, exp. 10/10/2013; 03712I000, exp. 10/10/13; when used with ARCHITECT Reaction Vessel (RV) list 7C15-01, box 2 lot numbers (bag 3 lot numbers): 19320P100 (6F9891501); 19344P100 (4N5891091); 19404P100 (1J1948151); 19322P100 (LM3891731); 19373P100 (6F9891521); 19407P100 (4N5948361); 19324P100 (NK8891921); 19374P100 (JG7891631); 19408P100 (5P4948471); 19340P100 (1J1890911); 19375P100 (LM3891741); 19497P100 (2L2948281); 19341P100 (2L2891001); 19379P100 (NK8891931); 20026P100 (LM3947841) - Amount recalled
- 28,600 total
Where it was sold
- Nationwide
Why it was recalled
The Architect 25-OH Vitamin D assay has the potential to generate falsely elevated or falsely depressed results when used with certain lots of Architect Reaction Vessels.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Abbott Laboratories recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.