Reported Jan 30, 2013 · Terminated

GreenLight HPS (High Performance System) GreenLight XPS (Xcelerated Performance System) Manufactured by AMS, Solutions…

AMS Innovative Center - San Jose

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: AMS Innovative Center - San Jose · 3070 Orchard Dr · San Jose, CA · 95134-2011

Is this the product you have?

  • GreenLight HPS (High Performance System) GreenLight XPS (Xcelerated Performance System) Manufactured by AMS, Solutions for Life Innovation Center. GreenLight HPS and XPS Surgical Lasers Systems and accessories are intended for use in endoscopic 532 nm laser resection of the prostate for the treatment of benign prostatic hyperplasia.
Lot codes and dates
  • HPS Part numbers0010-0070 to 0010-0077
  • XPS Part numbers0010-0210
  • Serial numbers

    389 of 389 codes

    • HPS2095R1
    • HPS3079C2
    • HPS3080C1
    • HPS3089C1
    • HPS3091C1
    • HPS3096C1
    • HPS3099C1
    • HPS3100C1
    • HPS3101C1
    • HPS3102C1
    • HPS3103
    • HPS3104C1
    • HPS3105C1
    • HPS3107
    • HPS3108C1
    • HPS3109C1
    • HPS3110C2
    • HPS3112
    • HPS3113
    • HPS3114
    • HPS3115
    • HPS3116
    • HPS3117
    • HPS3118
    • HPS3119
    • HPS3120
    • HPS3121
    • HPS3122
    • HPS3123
    • HPS3124
    • HPS3125
    • HPS3126
    • HPS3127
    • HPS3128
    • HPS3129
    • HPS3130
    • HPS3131
    • HPS3132
    • HPS3134
    • HPS3136
    • HPS3137
    • HPS3139
    • HPS3142
    • HPS3143
    • HPS3144
    • HPS3145
    • XPS50084
    • XPS50241
    • XPS50301
    • XPS50302
    • XPS50312
    • XPS50313
    • XPS50314
    • XPS50315
    • XPS50316
    • XPS50317
    • XPS50318
    • XPS50319
    • XPS50320
    • XPS50321
Amount recalled
389 devices, all model numbers.

Why it was recalled

During electrical recertification testing it was found that the Voltage Scaling Board was non-conforming to the IEC 60601-1-2 standard.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new AMS Innovative Center - San Jose recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 13, 2014 · Data as of Sep 19, 5:28 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.