Reported Dec 30, 2020 · Ongoing

Appx 1.7 ml Bag Spike w/Integrated Clave, 2 Drop-In Spiros, REF: CH3260

ICU Medical, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: ICU Medical, Inc. · 951 Calle Amanecer · San Clemente, CA · 92673-6212

Is this the product you have?

    • Appx 1.7 ml Bag Spike w/Integrated Clave
    • 2 Drop-In Spiros
  • REFCH3260
Lot codes and dates
  • Lots
    • 4741955
    • 4764030
Amount recalled
900

This recall covers 13 products

The agency filed one notice per product. You are reading one of them; the other 12 are below.

Where it was sold

Why it was recalled

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new ICU Medical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:54 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.