Reported Mar 7, 2018 · Terminated

AIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO¿ is intended to be used for X-ray computed…

Mobius Imaging, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Mobius Imaging, LLC · 2 Shaker Rd Ste F100 · Shirley, MA · 01464-2535

Is this the product you have?

  • AIRO Mobile CT System
  • Model NumberMobiCT-32
  • Product UsageThe AIRO¿ is intended to be used for X-ray computed tomography applications for anatomy that can be imaged in the 107cm aperture excluding pediatric patients and patients weighing over 400 lbs. (182 kg)
Lot codes and dates
  • Requested.
Amount recalled
105

Where it was sold

Why it was recalled

If the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing the axis at startup or preparing for a scan.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Mobius Imaging, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Aug 16, 2018 · Data as of Sep 19, 6:06 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.