Reported Dec 3, 2025 · Ongoing
Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
Philips Medical Systems Nederland B.v.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. · Veenpluis 4-6 · Best, N/A · N/A
Is this the product you have?
- Philips Azurion 7M20 systems with FlexArm ceiling-mounted system
- Model Number722234
- Lot codes and dates
- Model Number722234
- Amount recalled
- 172 units
Where it was sold
Why it was recalled
The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Medical Systems Nederland B.v. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.