Reported Jan 22, 2014 · Terminated
Merge Mammo is a multi-modality, vendor-neutral digital mammography workstation that enables imaging centers to use a…
GE Healthcare It
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: GE Healthcare It · 540 W Northwest Hwy · Barrington, IL · 60010-3051
Is this the product you have?
- Merge Mammo is a multi-modality, vendor-neutral digital mammography workstation that enables imaging centers to use a single workstation to display and read images from different vendors and different acquisition devices, including digital mammography, ultrasound, MRI and CT. Two and three dimensional image review, manipulation, analysis and therapy planning capabilities that support image management display needs in the medical environment from multiple locations within and outside the hospital. Productivity-Enhancing Second Console Workstations - Workstations designed to perform automated, routine tasks such as image review, printing and archiving as well as post processing capabilities that enable special services for referring physicians. Diagnostic Review Workstations - Workstations designed to assist radiologists and surgeons in conducting primary diagnostic review and planning through flexible and interactive manipulation of multi-modal softcopy image including the use of prosthetic template overlays, and including mammography. Physician's Review Workstations - Workstations designed to give easy and economic access to multi-modality softcopy images in multiple locations within and outside the hospital. (e.g. teleconferencing, teleradiology etc.) Lossy compressed mammographic images must not be used for primary diagnostic interpretation unless approved for use in digital mammography. Display monitors used for primary diagnostic interpretation of mammographic images must be approved for use in digital mammography.
- Lot codes and dates
- Software version 8.0.2
- Amount recalled
- 28 units
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
Why it was recalled
There is a potential safety issue with the 8.0.2 version software of the Merge (Cedara) Mammo viewer. The Merge Mammo viewer can load and display images not in synch with the selected or highlighted images in the thumbnail navigator. This incorrect image display issue could result in an inaccurate reading on the Merge Mammo workstation. This could result in a missed intervention or mis-diagnosis.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new GE Healthcare It recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.