Reported Mar 23, 2022 · Ongoing

Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intended…

Abbott Laboratories, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Abbott Laboratories, Inc · 1921 Hurd Dr · Irving, TX · 75038-4313

Is this the product you have?

    • Alinity s System software version2.8.0
    • REF LN 06P16-01
    • IVD
  • OEI(01)00380740138479 The Alinity s System is intended for In Vitro diagnostic use only. The Alinity s System is a highthroughput, fully automated immunoassay analyzer designed to determine the presence of specific antigens and antibodies by using chemiluminescent immunoassay technology
Lot codes and dates
  • Serial
  • numbers AS1001 through AS1387 UDIs(01)00380740138479(21)AS1002 (01)00380740138479(21)AS1005 (01)00380740138479(21)AS1009 (01)00380740138479(21)AS1012 (01)00380740138479(21)AS1013 (01)00380740138479(21)AS1016 (01)00380740138479(21)AS1019 (01)00380740138479(21)AS1023 (01)00380740138479(21)AS1028 (01)00380740138479(21)AS1031 (01)00380740138479(21)AS1032 (01)00380740138479(21)AS1044 (01)00380740138479(21)AS1058 (01)00380740138479(21)AS1064 (01)00380740138479(21)AS1097 (01)00380740138479(21)AS1126 (01)00380740138479(21)AS1148 (01)00380740138479(21)AS1153 (01)00380740138479(21)AS1155 (01)00380740138479(21)AS1160 (01)00380740138479(21)AS1161 (01)00380740138479(21)AS1193 (01)00380740138479(21)AS1212 (01)00380740138479(21)AS1213 (01)00380740138479(21)AS1215 (01)00380740138479(21)AS1217 (01)00380740138479(21)AS1218 (01)00380740138479(21)AS1219 (01)00380740138479(21)AS1220 (01)00380740138479(21)AS1221 (01)00380740138479(21)AS1222 (01)00380740138479(21)AS1224 (01)00380740138479(21)AS1225 (01)00380740138479(21)AS1226 (01)00380740138479(21)AS1227 (01)00380740138479(21)AS1230 (01)00380740138479(21)AS1231 (01)00380740138479(21)AS1233 (01)00380740138479(21)AS1234 (01)00380740138479(21)AS1235 (01)00380740138479(21)AS1236 (01)00380740138479(21)AS1237 (01)00380740138479(21)AS1238 (01)00380740138479(21)AS1240 (01)00380740138479(21)AS1241 (01)00380740138479(21)AS1242 (01)00380740138479(21)AS1243 (01)00380740138479(21)AS1246 (01)00380740138479(21)AS1247 (01)00380740138479(21)AS1248 (01)00380740138479(21)AS1249 (01)00380740138479(21)AS1250 (01)00380740138479(21)AS1251 (01)00380740138479(21)AS1252 (01)00380740138479(21)AS1253 (01)00380740138479(21)AS1254 (01)00380740138479(21)AS1255 (01)00380740138479(21)AS1256 (01)00380740138479(21)AS1257 (01)00380740138479(21)AS1259 (01)00380740138479(21)AS1261 (01)00380740138479(21)AS1262 (01)00380740138479(21)AS1263 (01)00380740138479(21)AS1264 (01)00380740138479(21)AS1266 (01)00380740138479(21)AS1267 (01)00380740138479(21)AS1268 (01)00380740138479(21)AS1269 (01)00380740138479(21)AS1270 (01)00380740138479(21)AS1271 (01)00380740138479(21)AS1272 (01)00380740138479(21)AS1273 (01)00380740138479(21)AS1274 (01)00380740138479(21)AS1275 (01)00380740138479(21)AS1276 (01)00380740138479(21)AS1277 (01)00380740138479(21)AS1278 (01)00380740138479(21)AS1279 (01)00380740138479(21)AS1280 (01)00380740138479(21)AS1281 (01)00380740138479(21)AS1282 (01)00380740138479(21)AS1283 (01)00380740138479(21)AS1284 (01)00380740138479(21)AS1285 (01)00380740138479(21)AS1286 (01)00380740138479(21)AS1287 (01)00380740138479(21)AS1288 (01)00380740138479(21)AS1289 (01)00380740138479(21)AS1290 (01)00380740138479(21)AS1291 (01)00380740138479(21)AS1293 (01)00380740138479(21)AS1294 (01)00380740138479(21)AS1295 (01)00380740138479(21)AS1296 (01)00380740138479(21)AS1299 (01)00380740138479(21)AS1300 (01)00380740138479(21)AS1301 (01)00380740138479(21)AS1302 (01)00380740138479(21)AS1303 (01)00380740138479(21)AS1304 (01)00380740138479(21)AS1305 (01)00380740138479(21)AS1306 (01)00380740138479(21)AS1307 (01)00380740138479(21)AS1309 (01)00380740138479(21)AS1310 (01)00380740138479(21)AS1311 (01)00380740138479(21)AS1312 (01)00380740138479(21)AS1313 (01)00380740138479(21)AS1314 (01)00380740138479(21)AS1315 (01)00380740138479(21)AS1316 (01)00380740138479(21)AS1317 (01)00380740138479(21)AS1318 (01)00380740138479(21)AS1319 (01)00380740138479(21)AS1320 (01)00380740138479(21)AS1321 (01)00380740138479(21)AS1322 (01)00380740138479(21)AS1323 (01)00380740138479(21)AS1325 (01)00380740138479(21)AS1326 (01)00380740138479(21)AS1327 (01)00380740138479(21)AS1330 (01)00380740138479(21)AS1332 (01)00380740138479(21)AS1333 (01)00380740138479(21)AS1334 (01)00380740138479(21)AS1336 (01)00380740138479(21)AS1341 (01)00380740138479(21)AS1344 (01)00380740138479(21)AS1346 (01)00380740138479(21)AS1347 (01)00380740138479(21)AS1348 (01)00380740138479(21)AS1349 (01)00380740138479(21)AS1350 (01)00380740138479(21)AS1351 (01)00380740138479(21)AS1352 (01)00380740138479(21)AS1353 (01)00380740138479(21)AS1354 (01)00380740138479(21)AS1355 (01)00380740138479(21)AS1358 (01)00380740138479(21)AS1359 (01)00380740138479(21)AS1360 (01)00380740138479(21)AS1368
Amount recalled
306 units

Where it was sold

Why it was recalled

Software error associated with the immunoassay analyzer wash cycle which is using 1 mL of wash buffer instead of the intended 3 mL of wash buffer to wash the exterior of the probe.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Abbott Laboratories, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 4:52 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.