Reported Mar 23, 2022 · Ongoing
Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intended…
Abbott Laboratories, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Abbott Laboratories, Inc · 1921 Hurd Dr · Irving, TX · 75038-4313
Is this the product you have?
Alinity s System software version2.8.0REF LN 06P16-01IVD
- OEI(01)00380740138479 The Alinity s System is intended for In Vitro diagnostic use only. The Alinity s System is a highthroughput, fully automated immunoassay analyzer designed to determine the presence of specific antigens and antibodies by using chemiluminescent immunoassay technology
- Lot codes and dates
- Serial
- numbers AS1001 through AS1387 UDIs(01)00380740138479(21)AS1002 (01)00380740138479(21)AS1005 (01)00380740138479(21)AS1009 (01)00380740138479(21)AS1012 (01)00380740138479(21)AS1013 (01)00380740138479(21)AS1016 (01)00380740138479(21)AS1019 (01)00380740138479(21)AS1023 (01)00380740138479(21)AS1028 (01)00380740138479(21)AS1031 (01)00380740138479(21)AS1032 (01)00380740138479(21)AS1044 (01)00380740138479(21)AS1058 (01)00380740138479(21)AS1064 (01)00380740138479(21)AS1097 (01)00380740138479(21)AS1126 (01)00380740138479(21)AS1148 (01)00380740138479(21)AS1153 (01)00380740138479(21)AS1155 (01)00380740138479(21)AS1160 (01)00380740138479(21)AS1161 (01)00380740138479(21)AS1193 (01)00380740138479(21)AS1212 (01)00380740138479(21)AS1213 (01)00380740138479(21)AS1215 (01)00380740138479(21)AS1217 (01)00380740138479(21)AS1218 (01)00380740138479(21)AS1219 (01)00380740138479(21)AS1220 (01)00380740138479(21)AS1221 (01)00380740138479(21)AS1222 (01)00380740138479(21)AS1224 (01)00380740138479(21)AS1225 (01)00380740138479(21)AS1226 (01)00380740138479(21)AS1227 (01)00380740138479(21)AS1230 (01)00380740138479(21)AS1231 (01)00380740138479(21)AS1233 (01)00380740138479(21)AS1234 (01)00380740138479(21)AS1235 (01)00380740138479(21)AS1236 (01)00380740138479(21)AS1237 (01)00380740138479(21)AS1238 (01)00380740138479(21)AS1240 (01)00380740138479(21)AS1241 (01)00380740138479(21)AS1242 (01)00380740138479(21)AS1243 (01)00380740138479(21)AS1246 (01)00380740138479(21)AS1247 (01)00380740138479(21)AS1248 (01)00380740138479(21)AS1249 (01)00380740138479(21)AS1250 (01)00380740138479(21)AS1251 (01)00380740138479(21)AS1252 (01)00380740138479(21)AS1253 (01)00380740138479(21)AS1254 (01)00380740138479(21)AS1255 (01)00380740138479(21)AS1256 (01)00380740138479(21)AS1257 (01)00380740138479(21)AS1259 (01)00380740138479(21)AS1261 (01)00380740138479(21)AS1262 (01)00380740138479(21)AS1263 (01)00380740138479(21)AS1264 (01)00380740138479(21)AS1266 (01)00380740138479(21)AS1267 (01)00380740138479(21)AS1268 (01)00380740138479(21)AS1269 (01)00380740138479(21)AS1270 (01)00380740138479(21)AS1271 (01)00380740138479(21)AS1272 (01)00380740138479(21)AS1273 (01)00380740138479(21)AS1274 (01)00380740138479(21)AS1275 (01)00380740138479(21)AS1276 (01)00380740138479(21)AS1277 (01)00380740138479(21)AS1278 (01)00380740138479(21)AS1279 (01)00380740138479(21)AS1280 (01)00380740138479(21)AS1281 (01)00380740138479(21)AS1282 (01)00380740138479(21)AS1283 (01)00380740138479(21)AS1284 (01)00380740138479(21)AS1285 (01)00380740138479(21)AS1286 (01)00380740138479(21)AS1287 (01)00380740138479(21)AS1288 (01)00380740138479(21)AS1289 (01)00380740138479(21)AS1290 (01)00380740138479(21)AS1291 (01)00380740138479(21)AS1293 (01)00380740138479(21)AS1294 (01)00380740138479(21)AS1295 (01)00380740138479(21)AS1296 (01)00380740138479(21)AS1299 (01)00380740138479(21)AS1300 (01)00380740138479(21)AS1301 (01)00380740138479(21)AS1302 (01)00380740138479(21)AS1303 (01)00380740138479(21)AS1304 (01)00380740138479(21)AS1305 (01)00380740138479(21)AS1306 (01)00380740138479(21)AS1307 (01)00380740138479(21)AS1309 (01)00380740138479(21)AS1310 (01)00380740138479(21)AS1311 (01)00380740138479(21)AS1312 (01)00380740138479(21)AS1313 (01)00380740138479(21)AS1314 (01)00380740138479(21)AS1315 (01)00380740138479(21)AS1316 (01)00380740138479(21)AS1317 (01)00380740138479(21)AS1318 (01)00380740138479(21)AS1319 (01)00380740138479(21)AS1320 (01)00380740138479(21)AS1321 (01)00380740138479(21)AS1322 (01)00380740138479(21)AS1323 (01)00380740138479(21)AS1325 (01)00380740138479(21)AS1326 (01)00380740138479(21)AS1327 (01)00380740138479(21)AS1330 (01)00380740138479(21)AS1332 (01)00380740138479(21)AS1333 (01)00380740138479(21)AS1334 (01)00380740138479(21)AS1336 (01)00380740138479(21)AS1341 (01)00380740138479(21)AS1344 (01)00380740138479(21)AS1346 (01)00380740138479(21)AS1347 (01)00380740138479(21)AS1348 (01)00380740138479(21)AS1349 (01)00380740138479(21)AS1350 (01)00380740138479(21)AS1351 (01)00380740138479(21)AS1352 (01)00380740138479(21)AS1353 (01)00380740138479(21)AS1354 (01)00380740138479(21)AS1355 (01)00380740138479(21)AS1358 (01)00380740138479(21)AS1359 (01)00380740138479(21)AS1360 (01)00380740138479(21)AS1368
- Amount recalled
- 306 units
Where it was sold
Why it was recalled
Software error associated with the immunoassay analyzer wash cycle which is using 1 mL of wash buffer instead of the intended 3 mL of wash buffer to wash the exterior of the probe.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Abbott Laboratories, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.