Reported Jan 24, 2024 · Ongoing

Cardinal Health Suction Canister Liner, MEDI-VAC CRD 1000 mL Suction Canister Liner Kit with 6 mm 1.8 m Tube, REF…

Cardinal Health 200, LLC

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Cardinal Health 200, LLC · 3651 Birchwood Dr · Waukegan, IL · 60085-8337

Is this the product you have?

  • Cardinal Health Suction Canister Liner, MEDI-VAC CRD 1000 mL Suction Canister Liner Kit with 6 mm 1.8 m Tube, REF 65651-517
Lot codes and dates
  • 50630140152670 (case)
  • Lot Numbers
    • J308-625
    • J308-629
    • J308-637
    • J309-616
    • J309-634
    • J309-646
    • J309-647
    • J310-603
    • J310-618
Amount recalled
74515

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Why it was recalled

The above-listed lots of Cardinal Health Medi-Vac CRD" Suction Canister Liners 1L may experience loss of suction on low or low intermittent suction settings, due to a misalignment of the liner with the outer hard canister.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Cardinal Health 200, LLC recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:38 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.