Reported Jan 22, 2014 · Terminated
MCA5-1SD - 1MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate…
The Anspach Effort, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: The Anspach Effort, Inc. · 4500 Riverside Drive · Palm Beach Gardens, FL · 33410-4235
Is this the product you have?
- MCA5-1SD - 1MM Diamond Ball, 5.6CM. All lots are included in recall. Intended for use in cutting and shaping delicate bone material, primarily in otologic procedures such as cochleostomy.
- Lot codes and dates
- All lots are included in recall
- Amount recalled
- 6439 total
This recall covers 28 products
The agency filed one notice per product. You are reading one of them; the other 27 are below.
15 more are in the archive.
Where it was sold
- Nationwide
- CA
Why it was recalled
The Anspach Effort, Inc. is recalling all lots of the Anspach Micro Curved Attachment (MCA) Cutting Burrs and Dissection Tools. The recall was initiated after the indications for use were found to be broader than the validation of the MCA Cutting Burrs and its directions for use did not clearly state the correct intended use of the device. After a labeling correction, Anspach Effort, Inc. decide
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new The Anspach Effort, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.