Reported Jan 22, 2020 · Terminated
QuickDraw Venous Cannula, REF: QD22, UDI: 00690103182699; and REF: QD25, UDI: 00690103182705
Edwards Lifesciences, LLC
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Edwards Lifesciences, LLC · 1 Edwards Way · Irvine, CA · 92614-5688
Is this the product you have?
- QuickDraw Venous Cannula
- REFQD22
- UDI00690103182699
- and REFQD25
- UDI00690103182705
- Lot codes and dates
- All lots
- Amount recalled
- 27168
Where it was sold
Why it was recalled
If the venous cannula is used in an unintended manner (longer than 6 hours; particularly during extracorporeal membrane oxygenation [ECMO]) there is a risk that separation of the cannula from its connector can occur; which can result in significant blood loss.