Reported Feb 20, 2013 · Terminated

Mizuho Orthopedic Table, model 5855, sub assembly 5855-901 Orthopedic Table Spar.

Mizuho OSI

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Mizuho OSI · 30031 Ahern Ave · Union City, CA · 94587-1234

Is this the product you have?

    • Mizuho Orthopedic Table
    • model 5855
    • sub assembly 5855-901 Orthopedic Table Spar
Lot codes and dates
    • Table model 5855
    • sub assembly 5855-901 Orthopedic Table Spar
    • manufactured between 1996 and 2007
Amount recalled
869 devices

Where it was sold

Why it was recalled

Possible instability of Orthopedic Table Top. If the epoxy bond fails, the aluminum plug may migrate out of the carbon fiber tube. If this occurs, the Orthopedic Table Spar may need to be replaced to ensure stability of the Orthopedic Table Top.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Mizuho OSI recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 15, 2016 · Data as of Sep 19, 6:50 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.