Reported Feb 13, 2019 · Terminated

Naera Hospital Bassinet, Model No. 4402

Stryker Medical Division of Stryker Corporation

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Stryker Medical Division of Stryker Corporation · 3800 E Centre Ave · Portage, MI · 49002-5826

Is this the product you have?

  • Naera Hospital Bassinet, Model No. 4402
Lot codes and dates
  • 2,174 of 2,174 codes

    • 2016000600003
    • 2016000600004
    • 2016000600005
    • 2016000600006
    • 2016000600007
    • 2016000600008
    • 2016000600009
    • 2016000600010
    • 2016000600011
    • 2016000600012
    • 2016000600013
    • 2016000600014
    • 2016000600015
    • 2016000600016
    • 2016000600017
    • 2016000600018
    • 2016000600019
    • 2016000600020
    • 2016000600021
    • 2016000600022
    • 2016000600023
    • 2016000600024
    • 2016000600025
    • 2016000600026
    • 2016000600027
    • 2016000600028
    • 2016000600029
    • 2016000600030
    • 2016000600031
    • 2016000600032
    • 2016000600033
    • 2016000600034
    • 2016000600035
    • 2016000600036
    • 2016000600037
    • 2016000600038
    • 2016000600039
    • 2016000600040
    • 2016000600041
    • 2016000600042
    • 2016000600043
    • 2016000600044
    • 2016000600045
    • 2016000600046
    • 2016000600047
    • 2016000600048
    • 2016000600049
    • 2016000600050
    • 2016000600054
    • 2016000600059
    • 2016000600060
    • 2016000600061
    • 2016000600062
    • 2016000600063
    • 2016000600064
    • 2016000600065
    • 2016000600066
    • 2016000600067
    • 2016000600068
    • 2016000600069
Amount recalled
5915

Where it was sold

  • Nationwide

Why it was recalled

It was identified that the volume of air inside the mattress may expand in regions of high altitude which may result in an unintended gap between the plateau of the inflated patient surface and tub. A Health Hazard Evaluation was completed which identified the potential risk of patients moving out of position and rolling over on the mattress. While the highest possible severity of harm associated with this condition is death, there have been no reports of injury or harm. This condition has only presented in regions of high altitude; however, Stryker has chosen to replace all Naera mattresses in the field.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Stryker Medical Division recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 29, 2020 · Data as of Sep 19, 6:17 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.