Reported Jan 29, 2014 · Terminated
ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21 Br¿nsh¿j, Denmark A portable, automated analyzer that measures…
Radiometer America Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Radiometer America Inc · 810 Sharon Dr · Westlake, OH · 44145-1521
Is this the product you have?
- ABL90 FLEX Analyzer Radiometer Medical ApS, ¿kandevej 21 Br¿nsh¿j, Denmark A portable, automated analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood.
- Lot codes and dates
Lot R0375N009Model #393-090S/N 393-090R0027N001 - 393-090R0379N010
- Amount recalled
- Total 3414 (US 524, CA 31, ROW 2859)
Where it was sold
Why it was recalled
RADIOMETER become aware that if the ABL90 FLEX is tilted, this may potentially cause a negative bias on sodium (Na+) results.