Reported Jan 29, 2020 · Terminated
Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model…
Isopure Corp.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Isopure Corp · 11851 Plantside Dr · Louisville, KY · 40299-6328
Is this the product you have?
- Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3
- Lot codes and dates
- IsoMix Pro UDI(01)00852456007085; ISOMIX Serial Number IMP-838703 with Pump SN 180210D18
- Amount recalled
- 1
This recall covers 3 products
The agency filed one notice per product. You are reading one of them; the other 2 are below.
Where it was sold
Why it was recalled
There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Isopure Corp. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.