Reported Mar 2, 2016 · Terminated

Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.

Draeger Medical, Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Draeger Medical, Inc. · 3135 Quarry Rd · Telford, PA · 18969-1042

Is this the product you have?

  • Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.
Lot codes and dates
  • Catalog #s 8416400 and 8417400
Amount recalled
2553

Where it was sold

  • Nationwide
  • PR

Why it was recalled

Stop using this product now

The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Draeger Medical, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Apr 1, 2020 · Data as of Sep 19, 5:19 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.