Reported Mar 2, 2016 · Terminated
9131-001 Defibrillation Electrodes, Lot No. 141125-02 Used in Powerheart G3 9300A, 9300E, 9300P, 9390A, and 9390E AEDs…
Cardiac Science Corporation
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Cardiac Science Corporation · N7 W22025 Johnson Dr · Waukesha, WI · 53186-1856
Is this the product you have?
- 9131-001 Defibrillation Electrodes, Lot No. 141125-02 Used in Powerheart G3 9300A, 9300E, 9300P, 9390A, and 9390E AEDs Cardiac Science 9131 Defibrillation Electrodes are single use and intended to be used in conjunction with Cardiac Science automatic external defibrillators (AED) to monitor and deliver defibrillation energy to the patient. The electrodes are intended for short term use (<8 hours) and must be used before the expiration date listed on the packaging. The AED electrodes are used for emergency treatment of cardiac arrest patients over 8 years of age or greater than 55 pounds. The user assesses the patient s condition and confirms that the patient is unconscious, pulseless and is not breathing prior to applying the electrodes to the skin.
- Lot codes and dates
- Lot No. 141125-02
- Amount recalled
- 10,943
Where it was sold
Why it was recalled
Defibrillation electrodes may increase electrical impedance over time. If impedance becomes too high, the AED will fail the electrode self-test and it will not be Rescue Ready, the AED STATUS INDICATOR will be RED and the AED will beep. This is normal behavior when the impedance limit is exceeded. The high impedance electrodes require replacement.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Cardiac Science Corporation recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.