Reported Feb 7, 2024 · Ongoing

Custom procedural convenience kits and trays, Ophthalmic use, labeled as: a) EYE PACK, kit number LVEY27; b)…

American Contract Systems, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: American Contract Systems, Inc. · 7802 E Telecom Pkwy · Temple Terrace, FL · 33637-0928

Is this the product you have?

  • Custom procedural convenience kits and trays, Ophthalmic use
  • labeled asa) EYE PACK, kit number LVEY27; b) Laminectomy, kit number UIOL34AG
Lot codes and dates
    • a) LVEY27
    • UDI/DI 00191072135728
  • Lot Numbers
    • 840221
    • exp. 6/29/2023
    • 771221
    • exp. 8/17/2023
    • 715221
    • exp. 10/12/2023
    • 982231
    • exp. 1/18/2024
    • 917231
    • exp. 3/24/2023
    • b) UIOL34AG
    • UDI/DI 00191072191168
  • Lot Numbers
    • 912231
    • exp. 3/29/2024
    • 898231
    • exp. 4/12/2024
Amount recalled
940 units

This recall covers 18 products

The agency filed one notice per product. You are reading one of them; the other 17 are below.

5 more are in the archive.

Where it was sold

  • Nationwide

Why it was recalled

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new American Contract Systems, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:17 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.