Reported Jan 29, 2020 · Terminated
Soredex DIGORA Optime SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
Palodex Group Oy
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Palodex Group Oy · Nahkelantie 160 · Tuusula · N/A
Is this the product you have?
- Soredex DIGORA Optime SOLID STATE X-RAY IMAGER (FLAT PANEL/DIGITAL IMAGER)
- Lot codes and dates
- Serial NumberSJ1208232, SJ1208233, SJ1208234, SJ1208235, SJ1208236, SJ1208238, SJ1208248, SJ1208249, SJ1208251, SJ1208252, SJ1208253, SJ1208254, SJ1208255, SJ1208256, SJ1208257, SJ1208268, SJ1208269, SJ1208270, SJ1208296, SJ1208297, SJ1208298, SJ1208299, SJ1208300, SJ1208301, SJ1208303, SJ1208304, SJ1208305, SJ1208306, SJ1208307, SJ1208309, SJ1208409, SJ1208415, SJ1208416, SJ1208417, SJ1208418, SJ1208421, SJ1208431, SJ1208433, SJ1208434, SJ1208436, SJ1208437, SJ1208439, SJ1208440, SJ1208441, SJ1208442, SJ1208446, SJ1208447, SJ1208449, SJ1208451, SJ1208452, SJ1208453, SJ1208454, SJ1208455, SJ1208456, SJ1208457, SJ1208460, SJ1208461, SJ1208462, SJ1208465, SJ1208519, SJ1208523, SJ1208524, SJ1208533, SJ1208534, SJ1208535, SJ1208536, SJ1208537, SJ1208538, SJ1208556, SJ1208558, SJ1208559, SJ1208560, SJ1208561, SJ1208563, SJ1208571, SJ1208572, SJ1208573, SJ1208574, SJ1208575, SJ1208577, SJ1208578, SJ1208579, SJ1208580, SJ1208581, SJ1208583, SJ1208584, SJ1208585, SJ1208586, SJ1208588, SJ1208596, SJ1208597, SJ1208598, SJ1208599, SJ1208600, SJ1208601, SJ1208602, SJ1208603, SJ1208604, SJ1208605, SJ1208606, SJ1208607, SJ1208608, SJ1208609, SJ1208610, SJ1210176, SJ1210177, SJ1210178, SJ1210179, SJ1210180, SJ1210181, SJ1210184, SJ1210185, SJ1210187, SJ1210188, SJ1210189, SJ1210191, SJ1210192, SJ1210193, SJ1210194, SJ1210195, SJ1210197, SJ1210200, SJ1210201, SJ1210202, SJ1210204, SJ1210205, SJ1210206, SJ1210208, SJ1210209, SJ1210210, SJ1210213, SJ1210219, SJ1210281, SJ1210282, SJ1210283, SJ1210284, SJ1210285, SJ1210287, SJ1210288, SJ1210289, SJ1210291, SJ1210292, SJ1210293, SJ1210295, SJ1210297, SJ1210298, SJ1210300, SJ1210310, SJ1210312, SJ1210313, SJ1210314, SJ1210315, SJ1210316, SJ1210317, SJ1210318, SJ1210319, SJ1210320, SJ1210322, SJ1210323, SJ1210324, SJ1210326, SJ1210398, SJ1210400, SJ1210401, SJ1210402, SJ1210404, SJ1210416, SJ1210417, SJ1210497, SJ1210499, SJ1210500, SJ1210501, SJ1210502, SJ1210503, SJ1210504, SJ1210506, SJ1210508, SJ1210511, SJ1210512, SJ1210514, SJ1210515, SJ1210528, SJ1210529, SJ1210530, SJ1210534, SJ1210535, SJ1210536, SJ1210538, SJ1210540, SJ1210541, SJ1210542, SJ1211114, SJ1211117, SJ1211118, SJ1211121, SJ1211123, SJ1211127, SJ1211129, SJ1211130, SJ1211131, SJ1211132, SJ1211133, SJ1211134, SJ1211135, SJ1211137, SJ1211138, SJ1211224, SJ1211225, SJ1211226, SJ1211227, SJ1211228, SJ1211229, SJ1211230, SJ1211231, SJ1211232, SJ1211233, SJ1211234, SJ1211235, SJ1211236, SJ1211237, SJ1211238, SJ1211239, SJ1211240, SJ1211241, SJ1211242, SJ1211243, SJ1211244, SJ1211245, SJ1211246, SJ1211247, SJ1211248, SJ1211249, SJ1211250, SJ1211251, SJ1211252, SJ1211263, SJ1211341, SJ1211343, SJ1211345, SJ1211346, SJ1211347, SJ1211348, SJ1211349, SJ1211354, SJ1211355, SJ1211356, SJ1211358, SJ1211360, SJ1211378, SJ1211379, SJ1211382, SJ1311602, SJ1311603, SJ1311605, SJ1311606, SJ1311607, SJ1311608, SJ1311609, SJ1311610, SJ1311611, SJ1311612, SJ1311613, SJ1311614, SJ1311615, SJ1311616, SJ1311617, SJ1311766, SJ1311767, SJ1311768, SJ1311769, SJ1311770, SJ1311771, SJ1311772, SJ1311774, SJ1311776, SJ1311777, SJ1311778, SJ1311779, SJ1311780, SJ1311781, SJ1311783, SJ1312271, SJ1312481, SJ1313119, SJ1313121, SJ1313122, SJ1313124, SJ1313127, SJ1313220, SJ1313223, SJ1313224, SJ1313225, SJ1313228, SJ1313230, SJ1314362, SJ1314363, SJ1314368 PSU's having nine-digit serial numbers
- beginning with four digits as follows1143XXXXX 1221XXXXX 1224XXXXX and 1229XXXXX
- Amount recalled
- 297
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
Why it was recalled
Slow degradation of plastic insulation used in the PSU DC plug of the DIGORA Optime, Scan eXam, Express, GXPS-500 systems may result in short circuiting which may cause heating and melting. The melting may generate smoke and eventually cause the PSU to fail.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Palodex Group Oy recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.