Reported Feb 5, 2020 · Ongoing
EQ-400-RH-QZ and EQ-400-LH-BK products Product Usage: The device is a very broadband light source emitting incoherent…
Energetiq Technology Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Energetiq Technology Inc · 7 Constitution Way · Woburn, MA · 01801-1024
Is this the product you have?
- EQ-400-RH-QZ
- and EQ-400-LH-BK products Product UsageThe device is a very broadband light source emitting incoherent non-laser light from the Deep UV through the visible wavelengths
- Lot codes and dates
- The defect concerns a total of fifteen (15) of our EQ-400 products. manufactured between June 3, 2016 and September 26, 2019.
- Amount recalled
- 15
Why it was recalled
On September 27, 2019 Energetiq Technology (the manufacturer) discovered a problem with their EQ-400-RH-QZ and EQ-400-LH-BK products during routine manufacturing functional testing. It was noted that a safety interlock within the lamp does not function in a reliable manner to shut down the system if the laser collimator is removed from the lamp while the system is operating. This is a potential safety hazard for anyone working on or nearby the lamp.
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Energetiq Technology Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.