Reported Mar 6, 2013 · Terminated

Versafitcup Impacting Ring Model Number: 01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10.0135 (50mm),…

Medacta USA Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Medacta Usa Inc · 4725 Calle Quetzal Ste B · Camarillo, CA · 93012-8429

Is this the product you have?

  • Versafitcup Impacting Ring Model Number01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10.0135 (50mm), 01.26.10.0136 (52mm), 01.26.10.0137 (54mm), 01.26.10.0138 (56mm), 01.26.10.0139 (58mm), 01.26.10.0140 (60mm), 01.26.10.0141 (62mm), 01.26.10.0142 (64mm). Designed for cementless use in total hip arthroplasty
Lot codes and dates
  • Reference Number Lot Number Reference Number Lot Number 01.26.10.0133: 095920, 096050, 096050A; 01.26.10.0138: 095925, 096055, 096055A; 01.26.10.0134: 095921, 096051, 096051A; 01.26.10.0139: 095926, 096056, 096056A; 01.26.10.0135: 095922, 096052, 096052A; 01.26.10.0140: 095927, 096057, 096057A; 01.26.10.0136: 095923, 096053, 096053A; 01.26.10.0141: 095928, 096058, 096058A; 01.26.10.0137: 095924, 096054, 096054A; 01.26.10.0142: 095929, 096059, 096059A.
Amount recalled
293 units

Where it was sold

Why it was recalled

Medacta USA is recalling Versafitcup Impacting Ring because the impaction ring may not fit properly in the implant shell in all cases. This could result in surgical delay potentially increasing possible complications under anesthesia, having to ream up to a larger shell size leaving less bone than planned and repositiong/removal of the shell after initial impaction.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Medacta USA Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Feb 23, 2013 · Data as of Sep 19, 4:53 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.