Reported Dec 10, 2025 · Ongoing

Reagent Kits/ containing flow cells, components of NextSeq 550Dx Instrument

Illumina, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Illumina, Inc. · 5200 Illumina Way · San Diego, CA · 92122-4616

Is this the product you have?

Reagent Kits/ containing flow cells, components of NextSeq 550Dx Instrument

7 products · 40 lot codes

  1. 20031098NextSeq 550Dx High Output Reagent Kit v2.5 (75 cycles) IVD, REF: 20028870/NextSeq 550Dx HO FC Cart v2.5, 75 Cycles

    Flow Cell REF
    20031098

    24 recalled lot codes

    • 00816270020132A184824-3
    • A185016-3
    • A183951-3
    • A182816-3
    • A183339-3
    • A182533-3
    • A183950-3
    • A184531-3
    • A182413-3
    • A184058-3
    • A183613-3
    • A183211-3
    • A183612-3
    • A183826-3
    • A182302-3
    • A183449-3
    • A182298-3
    • A182532-3
    • A182299-3
    • A183830-3
    • A183118-3
    • A182301-3
    • A184359-3
    • A184727-3)
  2. 20026365NextSeq 550Dx High Output Reagent Kit v2.5 (300 cycles) IVD, REF: 20028871/NextSeq 550Dx HO FC Cart v2.5, 300 Cycles

    Flow Cell REF
    20026365

    4 recalled lot codes

    • 00816270020118A182717-3
    • A182051-3
    • A179804-3
    • A183343-3)
  3. 20062316CN NextSeq 550Dx High Output Reagent Kit v2.5 (75 cycles) IVD, REF: 20064341/CN NEXTSEQ 550Dx HO FLOW CELL V2.5

    Flow Cell REF
    20062316

    2 recalled lot codes

    • 006975709330014A183402-4
    • A183550-4)
  4. 20062316CN NextSeq 550Dx High Output Reagent Kit v2.5 (300 cycles) IVD, REF: 20064344/CN NEXTSEQ 550Dx HO FLOW CELL V2.5

    Flow Cell REF
    20062316

    2 recalled lot codes

    • 006975709330014A183402-4
    • A183550-4)

Page 1 of 2

Amount recalled
11916

Where it was sold

Why it was recalled

An instrument intended for sequencing DNA libraries with in vitro diagnostic assays is used with reagent kits including flow cells that have flow cell gaskets that have a quality issue, which may lead to flow cell leaks during runs, which could cause run failures, that may result in: No/insufficient data output, instrument fluidics clogs, toxic substance exposure to user.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Illumina, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 7:32 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.