Reported Apr 13, 2022 · Ongoing

ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of…

Abbott Laboratories

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Abbott Laboratories · 100 Abbott Park Rd Bldg Ap8b · Abbott Park, IL · 60064-3502

Is this the product you have?

  • ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem with STAT protocol capability when used for the quantitative determination of myoglobin in human serum and plasma.
Lot codes and dates
  • UDI(01)00380740003272(17)220531(10)166000
Amount recalled
402 kits

Where it was sold

  • Nationwide

Why it was recalled

This calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in the Instructions for Use (IFU).

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Abbott Laboratories recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:30 PM UTC