Reported Apr 13, 2022 · Ongoing

BenchMark ULTRA and DISCOVERY ULTRA Instruments

Ventana Medical Systems Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ventana Medical Systems Inc · 1910 E Innovation Park Dr · Oro Valley, AZ · 85755-1962

Is this the product you have?

  • BenchMark ULTRA and DISCOVERY ULTRA Instruments
Lot codes and dates
  • Serial Numbers Less-Than-Or-Equal-To 322030
Amount recalled
9214

Where it was sold

Why it was recalled

Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.