Reported Jan 4, 2017 · Terminated

Specimen Gate Laboratory 5002-0180, Specimen Gate eReports 5002-0370, Software Version 1.0 Specimen Gate Laboratory is…

PerkinElmer Health Sciences, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: PerkinElmer Health Sciences, Inc. · 940 Winter St · Waltham, MA · 02451-1457

Is this the product you have?

  • Specimen Gate Laboratory 5002-0180, Specimen Gate eReports 5002-0370, Software Version 1.0 Specimen Gate Laboratory is intended for use as a data processing software used in the storage, retrieving, and processing of laboratory data. Specimen Gate eReports Software is a module of Specimen Gate Laboratory that functions to support newborn screening laboratories, hospitals, clinical and physicians in data management and generating reports
Lot codes and dates
  • Software Version 1.0
Amount recalled
21

Where it was sold

Why it was recalled

Issue may cause eReports to consume an excessive amount of disk space causing the software to stop operating.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new PerkinElmer Health Sciences, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated May 1, 2017 · Data as of Sep 19, 5:21 PM UTC