Reported Dec 24, 2025 · Ongoing

GE Healthcare Carestation anesthesia system, product number and REF numbers: Carestation 750 A1, REF 1012-9750-000;…

GE Medical Systems China Co., Ltd.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: GE Medical Systems China Co., Ltd. · Dev. Zone · Wuxi · N/A

Is this the product you have?

4 of 4 products

Products covered by this recall
Model / numberProduct as published
1012-9750-000GE Healthcare Carestation anesthesia system, product number and REF numbers: Carestation 750 A1
1012-9750-002Carestation 750 A2
1012-9755-000Carestation 750c A1
1012-9755-002Products for which affected PMB can be used as a spare part: Carestation 750c A2
Lot codes and dates
  • REF 1012-9750-000GTIN 00840682145596
  • REF 1012-9750-002GTIN 00840682146470
  • REF 1012-9755-000GTIN 00840682146425
  • REF 1012-9755-002GTIN 00840682146463
Amount recalled
888 units

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

  • Nationwide

Why it was recalled

Stop using this product now

Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new GE Medical Systems China Co., Ltd. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 6:22 PM UTC