Reported Jan 22, 2025 · Ongoing

Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N

Sin Sistema de Implante Nacional S.a

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: SIN SISTEMA DE IMPLANTE NACIONAL S.A · Jardim Analia Franco · Sao Paulo · N/A

Is this the product you have?

    • Implant Epikut S Plus CM 16-Degrees 4
    • 5 X 18
    • 0 mm
  • REFILM 4518N
Lot codes and dates
  • UDI-DI07899995273437
  • LotX010469757
  • Expiration2028/01
Amount recalled
123 (US 98, OUS 25)

Where it was sold

Why it was recalled

Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Sin Sistema de Implante Nacional S.a recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Data as of Sep 19, 5:44 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.