Reported Jan 14, 2015 · Terminated
IR remote control JUC, BUTTON, Material no.1239824, Serial number: to (incl.) 100139854. Patient positioning during…
Trumpf Medical Systems, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Trumpf Medical Systems, Inc. · 1046 Legrand Blvd · Charleston, SC · 29492-7672
Is this the product you have?
- IR remote control JUC, BUTTON, Material no.1239824
- Serial numberto (incl.) 100139854. Patient positioning during surgery, ranging from anesthesia induction to actual surgery to recovery from anesthesia. Patient transport on the operating tabletop, from the patient transfer system to the operating theatre or from the operating theatre to the patient transfer system
- Lot codes and dates
9 of 9 products
Products covered by this recall # Product as published 1 Mobile column JUPITER System U, Material no. 4500113 2 Mobile column JUPITER System Miyabi U, Material no. 1574977 3 Operating table column JUPITER SM 360 U, Material no. 1389824 4 Operating table column JUPITER SF U OCE, Material no. 1622774 5 Stationary column JUPITER System U, Material no. 4500112 6 FLOOR MOUNTING COLUMN JUPITER U, Material no. 1276277 7 Floor mounting column JUPITER 360 U, Material no.1389826 8 Operating table JUPITER UNIVERSAL U, Material no. 4500170 9 OR-TABLE JUPITER UNIVERSAL CARBON U, Material no.1223188. - Amount recalled
- 3303 total all devices
This recall covers 12 products
The agency filed one notice per product. You are reading one of them; the other 11 are below.
Where it was sold
Why it was recalled
Complaints have been reported of unintended movement occurring during surgical cases due to the use of damaged JUPITER remotes
This product could cause temporary or reversible health problems — stop using it.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Trumpf Medical Systems, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.