Reported Apr 20, 2022 · Terminated
Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384
Breas Medical, Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Breas Medical, Inc. · 16 Esquire Rd · North Billerica, MA · 01862-2527
Is this the product you have?
- Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384
- Lot codes and dates
Trolleys delivered between June 11th2020and December 3rd2021
- UDI07321820073849
- Amount recalled
- 707 units
Where it was sold
- Nationwide
Why it was recalled
Ventilator mounting brackets for the Vivo 50/60/65 and Vivo 45LS ventilator models can become detached from the Breas Ventilator Trolley, this may cause the ventilator to fall from the trolley and cause injury to a person who is in the immediate proximity
This product could cause temporary or reversible health problems — stop using it.