Reported Mar 6, 2019 · Terminated

AF 550 x 8ML/HR 2ML 15LO CPNBMT8015XL CP, Product code MT8015XL CPNB, Arrow AutoFuser Disposable Pain Control Pump

Arrow International Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Arrow International Inc · 2400 Bernville Rd · Reading, PA · 19605-9607

Is this the product you have?

    • AF 550 x 8ML/HR 2ML 15LO CPNBMT8015XL CP
    • Product code MT8015XL CPNB
    • Arrow AutoFuser Disposable Pain Control Pump
Lot codes and dates
  • 74B1602449
Amount recalled
1

This recall covers 17 products

The agency filed one notice per product. You are reading one of them; the other 16 are below.

4 more are in the archive.

Where it was sold

  • Nationwide

Why it was recalled

Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new infant & baby recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Nov 12, 2021 · Data as of Sep 19, 6:56 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.