Reported Feb 26, 2020 · Terminated
Trilogy EVO Ventilator, 02 - EU - Model Number: EU2100X15B Trilogy Evo is intended for pediatric through adult patients…
Philips Respironics, Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Philips Respironics, Inc. · 1001 Murry Ridge Ln · Murrysville, PA · 15668-8517
Is this the product you have?
- Trilogy EVO Ventilator, 02 -
- EU - Model NumberEU2100X15B Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories
- Lot codes and dates
- Model/Material NumberEU2100X15B
- Serial Numbers
H23987675BB18H243715497B08H243715587359H2476245817A0H255250576184H2552506568FEH25538501C37CH25538516AE1BH25538554EA69H255386102449H25538612075BH255386332562H2553865152A0H256064131C99H256064175ABDH2560642115E3H2560642336F1H25781710C84CH2578205663D2H257821415469H2578217429ACH25787398A23FH257930893CA2H255384784969H25538507A64AH247624629292H257872183623
- Amount recalled
- 27
This recall covers 13 products
The agency filed one notice per product. You are reading one of them; the other 12 are below.
Where it was sold
Why it was recalled
Stop using this product now
Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Philips Respironics, Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.