Reported Feb 10, 2021 · Terminated

Ethicon Dermabond Advanced Topical Skin Adhesive-topical skin adhesive to hold closed easily approximated skin edges of…

Ethicon, Inc.

Moderate riskThis product could cause temporary or reversible health problems — stop using it.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: Ethicon, Inc. · Us Highway 22 West · Somerville, NJ · 08876

Is this the product you have?

  • Ethicon Dermabond Advanced Topical Skin Adhesive-topical skin adhesive to hold closed easily approximated skin edges of wounds from surgical incisions
  • including incisions Product CodeDNX12
Lot codes and dates
  • Lot NumberQEBBTQ Exp
  • Date
    • April 30
    • 2022
  • GTIN10705031203532
Amount recalled
23,580 eaches

Where it was sold

  • Nationwide

Why it was recalled

Product distributed with a red tape as part of the sterile packaging which could impact the seal and result in a sterility breach

This product could cause temporary or reversible health problems — stop using it.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new Ethicon, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Nov 29, 2023 · Data as of Sep 19, 7:00 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.