Reported Feb 19, 2020 · Ongoing
Medtronic MiniMed 640G Insulin Infusion Pump, Ref#s/Model(s) - MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1751, and…
Medtronic Inc.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Medtronic Inc. · 18000 Devonshire St · Northridge, CA · 91325-1219
Is this the product you have?
Medtronic MiniMed 640G Insulin Infusion PumpRef#s/Model(s) - MMT-1511MMT-1711MMT-1512MMT-1712MMT-1751and MMT-1752 *Not for US Distribution
- Lot codes and dates
- All Serial Numbers ( excluding devices with black retainer rings) ***NOT DISTRIBUTED WITHIN THE US***
This recall covers 4 products
The agency filed one notice per product. You are reading one of them; the other 3 are below.
Where it was sold
Why it was recalled
Stop using this product now
There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Medtronic Inc. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.