GE Healthcare Lullaby Incubator Model Hood Door FRU kit, Part Numbers: M1169767, M1160674, M1187512, M1206646, M1166822
Wipro GE Healthcare Private Ltd.
High riskUsing this product could cause serious injury, illness or death — stop using it now.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Wipro GE Healthcare Private Ltd. · 122 Part 1 · Bengaluru, N/A · N/A
Is this the product you have?
- GE Healthcare Lullaby Incubator Model Hood Door FRU kit
- Part Numbers
M1169767M1160674M1187512M1206646M1166822
- Lot codes and dates
- ALL SERIAL NUMBERS
- Amount recalled
- 17 units
This recall covers 6 products
The agency filed one notice per product. You are reading one of them; the other 5 are below.
Why it was recalled
Stop using this product now
GE HealthCare has become aware that the bedside panel of Care Plus, Care Plus¿ Model 1000, 2000, 3000 and 4000 Incubator, Lullaby Incubator, Lullaby TR Incubator, and Lullaby XP Incubator can be upright and look closed but not be latched if the red indicator is visible. The portholes also can look closed when not latched when the portholes are obstructed e.g., hood covers. If a patient impacts the bedside panel or porthole that is not latched, the panel can disengage and fall open, no longer protecting the patient from falling.
Using this product could cause serious injury, illness or death — stop using it now.
Want the full recall guide?
The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.
Free recall alerts
Get alerted about new Wipro GE Healthcare Private Ltd. recalls
One email when a matching recall is published. No account, no cost, unsubscribe in one click.
We email you about recalls and nothing else. Every send has a one-click unsubscribe, and you can change your preferences any time without an account.