Reported Feb 12, 2020 · Terminated
Sof-Flex Double Pigtail Ureteral Stent Set, Global Product No. G15076 G14906 G14840 G14865 G17128
Cook Inc.
Moderate riskThis product could cause temporary or reversible health problems — stop using it.What you should do
Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.
Firm contact: Cook Inc. · 750 N Daniels Way · Bloomington, IN · 47404-9120
Is this the product you have?
- Sof-Flex Double Pigtail Ureteral Stent Set, Global Product No. G15076 G14906 G14840 G14865 G17128
- Lot codes and dates
NS7949093NS8077953NS8077954NS8077955807069181833428183343807069381721348187189NS807069284243047990841
- Amount recalled
- 13388 total
This recall covers 20 products
The agency filed one notice per product. You are reading one of them; the other 19 are below.
7 more are in the archive.
Where it was sold
- Nationwide
- PR
Why it was recalled
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.