Reported Apr 10, 2019 · Terminated

e600 Automatic Transport Ventilator, Model Number 01EVE600

O-Two Medical Technologies, Inc.

High riskUsing this product could cause serious injury, illness or death — stop using it now.

What you should do

Stop using the affected lot and contact the recalling firm for the return or refund process. Check the FDA enforcement report for lot-level detail.

Firm contact: O-Two Medical Technologies, Inc. · 45A Armthorpe Road · Brampton, Ontario · N/A

Is this the product you have?

    • e600 Automatic Transport Ventilator
    • Model Number 01EVE600
Lot codes and dates
  • EV60001 to EV60126
Amount recalled
119

This recall covers 3 products

The agency filed one notice per product. You are reading one of them; the other 2 are below.

Where it was sold

Why it was recalled

Stop using this product now

There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.

Using this product could cause serious injury, illness or death — stop using it now.

Want the full recall guide?

The guide adds the identifying codes laid out for scanning, symptoms to watch for, a timeline, and news coverage of this recall.

Free recall alerts

Get alerted about new O-Two Medical Technologies, Inc. recalls

One email when a matching recall is published. No account, no cost, unsubscribe in one click.

Where this came from

Source: U.S. Food & Drug Administration (openFDA)

Read the FDA (openFDA device enforcement) notice

Notice updated Jun 29, 2023 · Data as of Sep 19, 7:35 PM UTC

RecallNow is not a government agency. Every fact on this page is published by the FDA, USDA or CPSC and links back to their notice.